AAS
AA-7003 Automatic Flame/Graphite Furnace Atomic Absorption Spectrometer
- Increased stability from the suspension system of the optical system. Mild external influences and ambient temperatures have no effect on the instrument’s stability.
- One of the most compact AAS in the world. The AA-7000 series combines the flame atomic absorption spectrometer, graphic furnace atomic absorption spectrometer, optical system, atomizer, graphic furnace power supply, and electronics all into one unit.
- Long lasting nebulizers and burning heads. Using aero-space technology, pure industrial titanium nebulizers and burning heads are casted using a paraffin method. As such, they resist corrosion and oxidation and are capable of withstanding high temperatures.
- 1800 grooves/mm diffraction grating for higher resolution and higher energy
- Deuterium-lamp and background correction increase the accuracy of calibration results
- Reliable safety and alarm devices for lamps, graphite furnance, and atomizer
- Optional HG-01 hydride generator to analyze elements, such as As, Pb, Se, and Pb.
Features
- · PC controlled instrument and data-processing
- · Automatic wavelength positioning, slit switching, and lamp current and gain setup. all of which can be completed in under 40 seconds.
- · PC controlled six-lamp rotating tower which can automatically select element lamp to analyze up to 6 elements in one sample
- · Automatic fire ignition with gas flow and leakage automatic alarms
- · Optional nitrous oxide/C2H2 systems to analyze more than 30 elements that require high temperatures.
Technical Specifications
Spectrophotometric System
Wavelength Range: 190 – 900 nm
Spectral Bandwidth: 0.1, 0.2, 0.4, 1.0, 2.0 nm, with 5 second automated switching
Monochromator: C-T grating monochromator
Grating: 1800 / mm
Wavelength Accuracy: ± 0.2 nm
Wavelength Repeatability: better than 0.1 nm
Static Baseline Stability: ≤ ± 0.002 A/30min
Dynamic Baseline Stability: ≤ ± 0.004 A/30minBlazed Wavelength: 250 nm
Resolution: better than 0.3 nm
Background Correction: > 50 times
Flame Analysis
Cu Characteristic Concentration: ≤ 0.02 μg/ml
Detection Limit: ≤ 0.005 μg/ml
Precision: ≤ 1% RSDBurner Heads: 50 mm and 100 mm interchange titanium burner heads, titanium atomizer chamber
Position Adjustment: Adjusting mechanism for the best vertical positions and the rotation of the burner, takes one minute to switch between flame and hydride generation.
Graphite Furnace Analysis
Cd Characteristic Quantity: 0.5 × 10-12g
Detection Limit: 1.0 × 10-12g
Temperature: Room temperature – 3000 ℃
Temperature Programming: Up to 20 steps. Heating Modes: constant, steps, slope
Heating Rate: ≥ 2000 ℃ / secData Processing
Measurement Modes: flame absorption, flame emission, graphite furnace, hydride generation.
Analytical Methods: linear, nonlinear, standard addition method.
Printable Output: Calibration curve, the signal spectrum, instrument conditions, analysis parameters and analysis results (automatically stored and printed).Graphite Furnace (with Integrated Power Supply)
Dimensions (L x W x H): 880 mm x 540 mm x 450 (H) mmWeight: 125kg
Power: ~220 V, 50 Hz. Main Instrument: 200 W, Graphite furnace: 4 KWOur AAS Complete Set
1. Main Unit (selectable)
2. Computer, printer, specialized software
3. Gas source: air compressor, high-purity argon, high-purity acetylene gas (preparation required)
4. 4 Element lamps (based on selection)
5. Hydride generator (optional)
6. Graphite furnace auto-sampler (optional)
7. Graphite furnace cooling system (optional)
8. Spare parts and specialized tools